Double-access oven
Whether for use in the field of electronics, medicine, breeding, pharmacy or any other field requiring that the transport of samples be carried out as far as possible without contamination, the The double-access oven has many advantages. Thanks to its easy-to-clean interior, its rapid recovery times after opening the door or even the homogeneous temperature distribution, it represents the ideal material introduction device between a gray or black room and a clean room. The UF TS dual-access oven is based on the standard UFplus oven from Memmert and features all the latest technological features, such as product-specific heating mode and perfectly matched control technology. The TwinDISPLAY version guarantees precision work and optimal thermal safety in clean rooms.
Temperature range up to +250°C
4 model sizes (internal volumes from 161 to 749 liters)
1 version: TwinDISPLAY
Forced air circulation F
Model | Int. dim ( W x H x D ) mm | Temperature | Capacity | Technical sheet | Price | Cart |
---|---|---|---|---|---|---|
UF160TS | 745x1233x582 mm | T° amb + 10 à +250°C | 161 |
|
|
|
UF260TS | 825x1314x682 mm | T° amb + 10 à +250°C | 256 |
|
|
|
UF450TS | 1224x1233x782 mm | T° amb + 10 à +250°C | 449 |
|
|
|
UF750TS | 1224x1714x782 mm | T° amb + 10 à +250°C | 749 |
|
|
|
Int. dim ( W x H x D ) mm
Temperature
Capacity
Technical sheet
Int. dim ( W x H x D ) mm
Temperature
Capacity
Technical sheet
Int. dim ( W x H x D ) mm
Temperature
Capacity
Technical sheet
Int. dim ( W x H x D ) mm
Temperature
Capacity
Technical sheet
User-friendly and intuitive use
Electronic adjustment of air change rate and air damper position
Programming temperature/air change ramps for TwinDISPLAY models using AtmoCONTROL software
Preheating fresh air to prevent temperature fluctuations
Inner casing and outer wall made almost exclusively from high quality stainless steel, corrosion resistant and easy to maintain
Heating concept specifically adapted to the product for precise and homogeneous thermal control
Extensive programming and documentation options using interfaces, built-in status logging functionality and AtmoCONTROL software
3 years warranty (worldwide)
Temperature limitation. : +60, +70, +80, +95, +100, +120, +160, +180, +200 or +250 °C (to be specified when ordering) (A6)
Transparent glass door (4-layer insulating glass) Max temperature +250 °C (B0)
Glass door (4-layer insulating glass, borosilicate) Max temperature +300 °C (B1)
Interior lighting to allow observation of loading (R0)
Indoor socket (compatible only with temperature restricted to max. +60°C) (R3)
Virtually gas-tight interior casing (K2)
Passage of 23 mm internal diameter for the introduction of cables (F0)
Passage of 38 mm internal diameter, closable by flap, (F7)
Passage of 57 mm internal diameter, closable by flap, (F8)
Passage of 100 mm internal diameter, closable by valve, (F9)
4 to 20 mA current interface (0 to +310 °C temperature range) (V3)
Current interface 4 to 20 mA (temperature range 0 to +310 °C), temperature on internally flexible Pt100 probe for external temperature monitoring (1 SingleDISPLAY device maximum; 3 TwinDISPLAY devices maximum) ( V6)
Monitoring of fan speed with heating cut-off and alarm triggering in the event of an anomaly, option only valid for UFplus (V4)
Keyed safety lock (safety lock) (B6)
Opening the door on the left (B8)
Potential-free contact (24 V/2 A) with socket for displaying error messages. (H6)
Programmable process-dependent door locking (only on TwinDISPLAY devices) (D4)
Warning in case of open door (only on TwinDISPLAY devices) (V5)
Portable Pt100 probe, to be freely placed in an interior box or spot measurement on a sample, with socket to be mounted on a panel (H8)
MobileALERT, sending an SMS in the event of a malfunction or triggering of the device alarm (Option H6 required) (C3)
Castor frame (in two parts), height 140 mm (R9)
Stainless steel grid, electropolished
Perforated stainless steel tray
Built-in stainless steel drip tray, 15 mm edge (likely to disrupt the distribution of ambient temperature), cannot be used with option K1
Maximum load per built-in drip tray [kg]
Free-standing stainless steel drip tray, 15 mm edge (likely to disrupt the distribution of ambient temperature), cannot be used with option K1
USB-Ethernet adapter
5m Ethernet cable for connection to computer
USB stick with user ID (user ID): User license on USB stick for a defined device to prevent unauthorized intervention by third parties. Device number to be specified for any subsequent order (only for TwinDISPLAY devices).
Height adjustable feet (4 units)
Accessories (4 pieces) for stacking devices of the same dimensions
Straight chimney extension for air exhaust (60.3 mm outer diameter, 57 mm inner diameter), connection to a pipe if applicable
Angled chimney extension for air exhaust (60.3 mm outer diameter, 57 mm inner diameter), connection to a pipe if necessary. Only available with U, I and S devices
FDA-compliant software “AtmoCONTROL FDA-Edition”. Meets the requirements for the use of electronic data blocks in memory and for electronic signatures, defined in the Guidelines 21 CFR part 11 of the US Food and Drug Administration (FDA). Basic license for controlling a device (compatible with TwinDISPLAY devices only). IQ/OQ document available in German and English (without surcharge)
Calibration in the Memmert factory for one or more freely selectable temperature values
DQ/IQ/OQ/PQ qualification